- Prior to sampling, load each clean sampler, first with a sampling (impregnated) filter, then with a prefilter, separating the filters with a spacer. Ensure that the configuration in which the filters are loaded
leads to the sampled air passing first through the pre-filter and then through the sampling filter.
Note 1: Three-piece polystyrene cassettes are appropriate, with the middle ring section of the
cassette acting as the spacer. Impregnated filters that are stored in a desiccator can be
prepared up to 14 days prior to their use for sampling.
Note 2: Silica gel sorbent tubes [7] can be used in lieu of impregnated filters, but each sorbent tube
must be preceded by a pre-filter. - Calibrate each personal sampling pump with a representative sampler in the line.
- Sample accurately at 2 L/min for a total sample size of 30 to 500 L. Avoid sampler overloading [6].
- Transport the samples to the laboratory in a manner that prevents contamination or damage to the
filters. - Submit at least three blank impregnated filters as field blanks for each set of samples collected per
day. Handle these in the same way as the field samples. - Ship all samples to the laboratory in accordance with established chain-of-custody procedures [10].
- The samples can be stored at room temperature for one week; for longer storage, refrigerate the
samples (4 °C). - Analyze samples between 4 days and 4 weeks of sample collection.
SAMPLE PREPARATION: - Prior to carrying out sample dissolution, store the sampling filter for at least four days (to avoid nitric
acid losses). - Remove sample cassettes from storage and bring them to room temperature. Discard the pre-filters.
NOTE: If desired, the pre-filters can be analyzed for determination of particulate chlorides, bromides
and nitrates. If this analysis is carried out, it is recommended to rinse the inside surfaces
of the sampler with a few mL of DI water so as to include wall deposits along with material
collected on the prefilter. - Place the sampling filters (i.e., impregnated filter samples) in 15-mL plastic screw-cap vessels and
add 10 mL of deionzied water to each sample. Securely cap the vessels. - Sonicate the samples in an ultrasonic bath for at least 15 minutes, and allow to cool for approx. 30
minutes. - Using 5-mL syringes, filter 5-mL aliquots of each sample extract solution through a PTFE filter into
clean auto-sampler vials.
CALIBRATION AND QUALITY CONTROL: - With dilution of the calibration stock solution in eluent solution, prepare calibration working
standard solutions covering the range of approximately 0.4 to 4 mg/L of chloride, bromide and
nitrate. Store working standards in tightly sealed polyethylene bottles. Prepare fresh working
standards weekly. - Calibrate the ion chromatograph with at least six working standards covering the range of 0.4 to
4 mg/L of chloride, bromide and nitrate per sample by preparing a calibration graph of anion peak
height (mm or µS [micro-siemens]) vs. concentration (mg/L).
NIOSH Manual of Analytical Methods (NMAM), Fifth Edition
VOLATILE ACIDS by Ion Chromatography: METHOD 7907, Issue 1, dated 20 May 2014 - Page 4 of 6 - Analyze working standards together with samples, reagent blanks and field blanks at a frequency of
at least 1 per 20 samples (3 minimum of each).
MEASUREMENT: - Set the ion chromatograph to recommended eluent flow rate (e.g., 1.5 mL/min) and recommended
pressure ( e.g., 1.1 x 105
kPa), and other conditions as specified by the instrument manufacturer. - Inject a sample aliquot, e.g., 25-µL, into the chromatograph, and measure the peak heights of
the chloride, bromide and nitrate peaks (at retention times of approx. 5 min, 9 min and 12 min,
respectively). If the peak height exceeds the linear calibration range, dilute with eluent, reanalyze
and apply the appropriate dilution factor in calculations.
CALCULATIONS: - Calculate the mass concentration of each anion, C (mg/m3
), in the air volume sampled, V (L):
C C V F C V
V
F d = c − {
( ) ( )} * 1* 1 0 * 0*
where:
C0
=mean concentration, in mg/L, of anion in the field blank test solutions;
C1
=concentration, in mg/L, of anion in the sample test solution;
V = volume, in liters, of the air sample;
V0
= volume, in mL, of the field blank test solutions;
V1
= volume, in mL, of the sample test solutions
Fd
= dilution factor for each sample test solution
Fc
= conversion factor to convert from anion to acid concentration
Fc
= 1.0284 for chloride, 1.0126 for bromide, and 1.0163 for nitrate
Sample Solution